MultiFlo™ FX is an automated reagent dispenser for 6- to 1536-well microplates. Its unique Parallel Dispense™ technology allows up to four independent reagents to be dispensed in parallel without potential carryover. What makes MultiFlo FX “multi-mode” is its configurability: you can add a wash module or the RAD™ module, for random access individual well dispensing and bulk dispensing larger wells. Everything you need for fast, intuitive programming and operation is available through the color touchscreen. MultiFlo FX configured with either the wash module or RAD technology replaces up to five liquid handlers in a single compact platform.
MFX | MFXP* | MFXW** | MFXPW** | MFXR*** | MFXPR*** | |
Unconfigured base instrument |
• |
|||||
1 peripump dispenser only |
• |
|||||
Wash module only |
• |
|||||
1 peripump dispenser and wash module |
• |
|||||
RAD technology with one peripump | • | |||||
RAD technology with two peripumps | • |
GENERAL | |
Microplate types | 96-, 384- and 1536-well |
Low profile and standard height | |
Solid and filter bottom (option) | |
Onboard software | Create, edit or run multiple protocols |
Software (computer control) | LHC2 Software |
LHC2 Secure for 21 CFR Part 11 compliance (option) | |
SiLA Compliant driver (option) | |
Separation | Biomagnetic separation, vacuum filtration (optional) |
Shaking | Programmable up to 60 minutes |
Slow, medium, fast or variable | |
Soaking | Programmable up to 60 minutes |
Ultrasonic Advantage™ | Standard on most configurations |
Automation | BioStack and 3rd party automation compatible |
BioSpa 8 Automated Incubator compatible | |
WASHING | |
Manifold types | 96-well washing: 96-tube manifold – straight dispense tubes |
96- and 384-well washing: 96-tube Dual-Action manifold – 20° angled dispense tubes | |
384-well washing (fast): 192-tube Dual-Action manifold – 7° angled dispense tubes | |
1536-well washing: Two 32-tube dispense manifolds – 7° angled dispense tubes | |
Volume range | 3 – 3,000 µL/well in 1 µL increments |
Wash cycles | 1 – 250 |
Buffer/reagent selection | Auto switching module for up to 4 buffers (option) |
Supply bottle | 4 L or 10 L (optional) |
Dispense accuracy | 3% |
Dispense precision | <3% CV (model dependent) |
Residual volume | <2 µL/well |
Wash speed | 96 wells, 300 µL/well, 96-tube manifold: 13 seconds |
384 wells, 100 µL/well, 192-tube manifold: 17 seconds | |
1536 wells, 10 µL/wells, two 32-tube manifolds: 36 seconds | |
Flow rates | High flow to low flow |
Optimized rates for cell assays | |
Sterilization | Chemical |
Vacuum range for filtration | Selectable: 0 to -380 mmHg (final at 30s) |
Filtration time: 5 – 999 seconds | |
DISPENSING – PERISTALTIC PUMP (MULTI-CHANNEL) | |
Manifold types | 8-tip (1 x 8) cassette with plastic, 316 SS tubes or sapphire jeweled 316 SS tubes |
Fluid delivery | 1 or 2 peristaltic pumps |
Dispense speed | 96 wells, 10 µL /well: 8 seconds |
384 wells, 5 µL /well: 12 seconds | |
1536 wells, 1 µL /well: 27 seconds | |
Volume range | 500 nL – 3,000 µL/well, selectable in 1 µL increments |
Flow rates | User programmable rates from high to low |
Optimized rates for cell assays | |
Dispense performance | 1 µL cassette: |
Recommended Volume Range: 1 – 50 µL | |
Dispense Accuracy: +5% at 1 µL | |
Dispense Precision: <5% CV at 1 µL | |
<10% CV at 500 nL | |
Minimum Prime Volume: 1.20 mL | |
5 µL cassette: | |
Recommended Volume Range: 5 – 2,500 µL | |
Dispense Accuracy: +2.0% at 5 µL | |
Dispense Precision: <2.5% CV at 5 µL | |
Minimum Prime Volume: 4.23 mL | |
10 µL cassette: | |
Recommended Volume Range: 10 – 3,000 µL | |
Dispense Accuracy: +2.0% at 10 µL | |
Dispense Precision: <2.0% CV at 10 µL | |
Minimum Prime Volume: 7.36 mL | |
Recommended cassette replacement interval | 1 µL Cassette: 1,000 384-well microplates at 5 µL well |
5 µL Cassette: 1,000 96-well microplates at 50 µL well | |
10 µL Cassette: 1,000 96-well microplates at 100 µL well | |
Sterilization | Autoclave, chemical |
DISPENSING – SYRINGE PUMP | |
Manifold types | 96-well dispensing: |
One 16-tube (2 x 8) manifold – 316 SS tubes | |
96-/384-well dispensing: | |
Two 16-tube (1 x 16) manifolds – 316 SS tubes | |
1536-well dispensing: | |
Two 32-tube (1 x 32) manifolds – sapphire jeweled 316 SS or 316 stainless steel tubes | |
Fluid delivery | Two positive displacement syringe drives |
Dispense speed | 20 µL /well, 96 wells, 1 x 16 tubes: 5.25 seconds |
20 µL /well, 384 wells, 1 x 16 tubes: 14 seconds | |
3 µL /well 1536 wells, 2 x 32 tubes: 7 seconds | |
Volume range | 3 – 3,000 µL/well, selectable in 1 µL increments |
Minimum Prime Volume: 12 mL | |
Flow rates | User programmable rates from high to low |
Dispense accuracy | +1 µL at 5 µL |
+1 µL at 20 µL | |
+1% at 100 µL | |
Dispense precision | <5% CV at 5 µL |
<2.5% CV at 20 µL | |
<1% CV at 100 µL | |
Supply bottle | 1 L or 2 L |
Sterilization | Chemical, autoclavable option |
PHYSICAL CHARACTERISTICS | |
Power | 100 – 240 Volts AC. 50/60 Hz |
Dimensions | 16.5″ W x 18″ D x 12.5″ H (42 x 46 x 32 cm) |
Weight | 32 lbs (14.5 kg) |
REGULATORY | |
Regulatory | CE and TUV marked. ROHS compliant. In Vitro Diagnostic use models are available. |